Specifications
Input type and range
J: –392 to 1832°F (–200 to 1000°C)
K: –418 to 2501°F (–250 to 1372°C)
Accuracy: ±0.8°F (±0.5°C), and ±0.1% of reading
above –238˚F (–150˚C), ±2°F (±1°C), and ±0.25% of
reading below –238˚F (–150˚C)
Resolution: 0.1°/1° selectable up to 999.9°;
1° above 1000°; autorange to 1° below –150°C
Display: 12-character alphanumeric LCD
Display update rate: 3 seconds/channel
Scan rate: from 4 seconds to 99 minutes, 59 seconds
Digital output: RS-232, ASCII format, baud rate 19.2
K, RJ11 connector; Centronics parallel printer output;
DB25F connector
Power: 115 or 230 VAC adapter (included);
10 to 28 VDC or 7 to 20 VAC 500 mA power supply
Dimensions: 10
1
2"W x 3
1
2"H x 7
1
2"D
Catalog number Description
WD-69200-00
12-channel scanning thermometer
12-Channel
Scanning Thermometer
Stores and recalls up to 4680 sets
of readings with time and date
Automatically or manually display,
record, and output temperature readings
Field calibrate each probe separately—
achieve accuracies of ±0.1% of reading!
Bidirectional RS-232 for setup
or PC datalogging
The 12-channel scanning thermometer is your best choice when simultaneous
monitoring of temperatures during different processes or phases is critical.
Portable unit accepts most any power source giving you great flexibility as
to where and when it may be used. Functions as a datalogger—store data
in nonvolatile memory and later download to your printer or PC. Link up to
eight of either model in series to expand the number of channels you can
scan for a maximum of 96 channels. Each of the 12 thermocouple inputs
are programmable to scan from every four seconds to every hour. Graph in
real time—use CD-ROM software (included). Includes a real-time clock and
nonvolatile memory. RF immunity—maintains accuracy in RF fields.
Parallel printer connection allows you to connect the thermometer to a
printer—make a hard copy of your data as a permanent record. Power unit
with included 115 VAC adapter or use with any 10 to 28 VDC or 7 to 20 VAC
500 mA power supply.
Instrument FeaturesScan and log up to 12 thermocouple inputs from
any combination of type J and K thermocouple probes. All thermocouple types
are individually selectable per channel. Accepts thermocouple probes with
miniconnectors for quick and easy probe changes. Records temperature data
sequentially so that process variations and thermal profiling can be monitored.
Datalogging feature automatically stores up to 4680 sets of readings and then
allows the data to be downloaded to a computer. Up to eight units can be
linked together for a maximum channel capacity of 96.
The easy-to-use front panel menu keys allow the user to program datalogging
and print intervals, scan rates, date and time, temperature scales (°F, C, K, or R),
resolution (0.1° or 1°), channels displayed, thermocouple type, and data output
parameters. Each channel features HOLD, STORE, MIN and MAX, RECALL, and
AVERAGE buttons. All channels can be calibrated in the field at one or two points
to ensure accuracy.
Multiple Safety FeaturesThermocouple probes are electrically isolated
from each other and data outputs to provide safe and accurate measurements
even when probes are exposed to harsh electrical conditions. Each channel
offers high and low audible alarm selectable set points.
Physical FeaturesEasy-to-read,
3
4" high, 14-segment, 12-character
alphanumeric LCD indicates channel number or name, temperature reading,
temperature scale, open probe conditions, and out-of-range readings.
Computer/Printer CapabilitiesRS-232 serial communication compatible.
Software allows data to be downloaded to a computer and then displayed in
real-time tabular or graphical format (software and cable included). Optional
parallel printer cable enables scanner to be connected directly to a printer,
without using a computer. Datalogging and printing intervals can be set from
four seconds to 99 minutes, 59 seconds.
What’s included: software and cable.
Benchtop 12-channel
scanning thermometer
Accessories
WD-22050-04 Parallel printer cable, 36M/DB25M
FREE Real-Time Logging Software
Unit includes FREE Windows
®
95/98NT/2000/XP-compatible software on
CD-ROM that controls and displays up to 12 scanners simultaneously.
Logged data may be saved to a file and imported into spreadsheet or
word processor. Live updating graph shows time and temperature!
View channel temperature—
push the “virtual buttons” to
change thermometer functions
Live, real-time graphs
auto-adjust the time on the Y-axis
Back view; accepts any type J or K
thermocouple probe with ANSI miniconnectors